High accuracy photometric ozone analyzer. This unit is ideal for laboratory or research applications where accuracy is a necessity.
| Reporting Units: |
ppm, ppb, mg/m3, micro g/m3 / ug/m3 |
| Zero noise: |
<0.3 ppb (RMS) |
| Span noise: |
<0.5% of reading (RMS) (above 100 ppb) |
| Lower detectable limit (LDL): |
<0.6 ppb (RMS) (2x noise as per US EPA) |
| Zero drift: |
<1 ppb/24 hours, <1 ppb/7 days |
| Span drift: |
<1% reading/7 days |
| Lag time: |
<10 seconds |
| Rise and fall time: |
<20 seconds to 95% |
| Linearity: |
1% full scale |
| Precision: |
0.5% of reading`above 50 ppb |
| Pneumatics: |
|
|
Internal diaphragm-type pump |
|
800 cm3/min +/- 10% |
|
47 mm diameter particulate filter, front-panel accessible |
| Operating Range: |
5-40 ºC (with EPA equivalency) |
| Dimensions (HxWxD): |
7" (178 mm) x 17" (432 mm) x 23.5" (597 mm) |
| Weight: |
28 lbs. (13 kg); 31 lbs. (14 kg) with IZS installed |
| Power: |
100-130 or 220-240 V, both at 50/60Hz; 250 W |
| Approvals: |
US EPA designation EQOA-0992-087 CE Mark Compliance |
| Outputs: |
Analog 0.1V/1V/5V/10V full scale, user selectable; -5 to +10% offset 8 digital status outputs, 6 inputs |
|
Digital: COM1: RS-232 Serial Port (DB-9M connector) (incl. cable)
COM2: RS-232 or RS-485 (DB-9F connector) |